Hirayama HVE-50 Top Load Sterilizer, JKKP approved in Malaysia
Hirayama HVE-50 Top Load Sterilizer AT-HVE-50: Self-Contained, Portable, Top-Loading Autoclaves with Auto Exhaust and Warming Cycles no dry without basket
Space-saving design with lid opening upward.
Agar sterilization mode for sterilizing agar and then automatically maintaining it at liquid temperature.
Dual-sensing lid interlock prevents lid opening when chamber is pressurized or when temperature exceeds 97°C.
Agar melting mode for melting agar and maintaining it at liquid temperature, or for a hot, faster start-up.
One-touch lever to easily open and close chamber lid.
In-process display shows the status and the progress of sterilization cycle.
Door-closure sensor provides a prompt and the cycle will not start unless the door is closed properly.
HVE-50 size: 50 liters, 11.8″ D, 28.0″H.
TEMPERATURE | |
Operating Range |
105ºC to 135ºC
|
Display Range |
5ºC to 137ºC
|
Warming Range |
45ºC to 60ºC
|
GAUGE PRESSURE | |
Operating Range (gauge) |
3 to 30 psi (0.021 to 0.206MPa)
|
Maximum Allowable |
38 psi ( 0.26MPa)
|
TIMER | |
Sterilization |
1 to 250 minutes
|
DIMENSIONS | |
Overall (W x D x H) |
20.1″ x 20.8″ x 40.9″ (51 cm x 53 cm x 104 cm)
|
Chamber (Diameter x Height) |
50 liters
11.8″ x 28″ (30 cm x 71 cm) |
SAFETY DEVICES |
Dual-sensing lid interlock, over-pressure power cutoff, over-temperature power cutoff, low-water power cutoff, mechanical pressure relief valve, lid-closure detector, over-current detector and circuit breaker
|
PROGRAMS | |
1. For agar/liquids. Also for melting agar, or for faster start-up |
Heating -> Sterilization -> Auto steam exhaust or ambient cooling (programmable) -> Warming -> Completion
|
2. For liquids/solids |
Heating -> Sterilization -> Auto steam exhaust or ambient cooling (programmable) -> Completion
|
3. For solids |
Heating -> Sterilization -> Auto exhaust (fast) -> Completion
|
ACCESSORIES | |
Stainless steel |
Mesh wire baskets, buckets (pails), and round test tube racks
|
Factory or field installed |
Data logger
|
POWER | |
Source |
120V, 50/60 Hz, single phase (220V available on request)
|
Consumption |
2000 watts (2000 VA)
|
SHIPPING WEIGHT |
180 lb (82 kg)
|
Sterilizer Type | Automatic Sterilizer |
Sterilizer Door Size | 12 inches |
Sterilizer Chamber Depth | 28 inches |
Sterilizer Dimension | 20″ x 21″ x 41″ (510mm x 530mm x 1040mm) |
Sterilizer Chamber Volume | 13 Gal/50 L |
Sterilizer Power Voltage | 120V Single Phase |
Sterilzier Power Current | 20A |
Dry Mode | No |
Loading Methods | Top Load (Vertical) |
Sterilizer Program | 1. For agar/liquids/for melting agar, or for faster start-up, 2. For liquid/solids, 3 For solids (Hirayama) |
Sterilizer High Temperature | 275F (134C) |
Color | Beige |
Type of Heating Element | Internal Heating Element |
Type of Door | Hinge Door |
Laboratory Customer | Yes |
Industry Customer | Yes |
Government Customer | Yes |
Tattoo, Beauty Salon Customer | No |
Medical Doctor | Yes |
Dentist Customer | Yes |
FDA Certificate | No |
ASME Certificate | No |
ISO Certificate | Yes |
Printer | Option |
Warranty Parts and Labor | 1 year |