TOMY Sliding Door Type Series ES-215/315 Autoclaves

Description

Autoclaves Sliding Door Type Series High Pressure Steam Sterilizer E-315/215

  • Provided with an Automatic Air Release Time Setting Function

    • ES-315
    • ES-215

 

Reliable & Speedy Sterilization

A top-open lid, which can be operated easily with a single hand or foot, is provided with the SX autoclaves. To open the lid, the lid lock is released by stepping on the foot pedal locates at the bottom while pressing the lid down. The lid can be opened with minimal effort owing to mechanical assistance. In addition, the lid can be closed easily just by pressing it down.

User-friendly Memory Function

Once turned on the system automatically sets and displays a pre-set value. Operation is done simply by pressing the start switch. ES-315/ES-215 enables the set temperature and time for each mode program to be stored in its memory.

Easy-to-see Control Panel*The control panel lighting image shown above may vary from actual display.

MODE/PROCESS display enables real time monitoring of operation status.

ES-315 / ES-215 offers three mode programs.

Sterilization Mode Sterilization Mode
It contains a delay start timer function that allows operation to start 1 to 99 hours later.

Sterilization/warming Mode Sterilization/warming Mode
This mode is useful when not removing culture media from the autoclave immediately after sterilization. Sterilized media can be kept warm to prevent coagulation.

Heating Mode Heating Mode
This mode program automates the dissolution of culture media that can sometimes be a painful process when using a water bath to boil them.

User-friendly design allows for even more easy maintenance.

Chamber having a flat inner surface

  • All models are supplied with stainless steel wire mesh baskets. No protrusions on inner surface of the chamber enable smooth loading and unloading of the basket as well as block design of all pipe fittings ensures safe operation.

Front access for easy maintenance

  • The exhaust bottle to capture and store exhaust steam is placed on the front panel of the autoclave unit. Access to the chamber for replacing water inside, as well as for routine maintenance and inspection, is done on the front of the unit, which also permits operators to replace bellows, saving costs for services by the service personnel.

A microprocessor controlled monitoring system provides safe sterilization.

Incomplete sterilization protection

  • In case of abnormal chamber temperature rise, the autoclave stops its operation. In case of abnormal decrease in temperature, the sterilization timer is temporarily suspended and then reactivated when the set temperature is reached.

Chamber overpressure protection

  • If overpressure in the chamber is detected, the autoclave stops all the function and indicates an error message simultaneously.

Chamber interior overtemperature protection

  • If the temperature inside the chamber exceeds the set temperature by 5°C or more, the autoclave stops all the function and indicates an error message simultaneously to avoid overheating.

Chamber exterior overtemperature protection

  • The thermostat switch detects the temperature of chamber exterior wall, which allows the autoclave to stop all the function if the temperature exceeds 130°C.

Specifications

Model ES-215 ES-315
Operating temperature range
(during sterilizing)
105 – 123°C 105 – 132°C
Operating pressure range 0 – 127kPa 0 – 186kPa
Maximum operating pressure 147kPa 216kPa
Temperature control Digital, microprocessor controlled
Temperature display / display range Digital / -15 – 180°C
Pressure display / display range Analog / 0 – 250kPa Analog / 0 – 400kPa
Heat source 1.5kW electric heater 2.0kW electric heater
Timer control Digital, microprocessor controlled
Timer display Digital
Timer
display range
Sterilizing mode: 1 – 240min. or continuous
Sterilizing-warming mode: 1 – 240min. (sterilizing) / 1 – 8hours (warming)
Heating mode: 1 – 8hours
Start-timer setting range 1 – 99hours in each operation mode
Memory function A set of temperature and time setting in each operation mode
Safety devices • Inside the chamber overheat prevention, • Outer wall of the chamber overheat prevention, • Overpressure prevention, • Temperature sensor disconnection prevention, • Empty heating prevention, • Leakage breaker, • Safety valve
Malfunction prevention devices • Low water level detection, • Exhaust valve knob open/close detection,
• Insufficient sterilization detection, • Chamber lid open/close detection
Leakage
breaker
Rated breaking current: 20A (120V) , 15A (220 / 230 / 240V) 30A (120V) , 15A (220 / 230 / 240V)
Rated sensed current: 30mA (120V) , 10mA (220 / 230 / 240V)
Protection type Class 1 equipment
Operating
environment
Ambient temperature: 5 – 35°C
Relative humidity: 30 – 85%
Atmospheric pressure: 700 – 1060hPa
Gradient: Within 2°
Dimensions φ248 x 543mm φ325 x 740mm
Usable dimension φ236 x 450mm φ312 x 635mm
Capacity of chamber 22ℓ 53ℓ
Chamber material SUS304
Capacity / material of exhasut bottle 3ℓ / polyethylene
Type / material of lid gasket Accommodating internal pressure / silicone rubber
Dimensions of main unit (mm) 400W x 460D x 920H (Height from floor to control panel:705) 490W x 560D x 1090H (Height from floor to control panel:875)
Net weight 50kg 80kg
Rated Voltage AC 120 / 220 / 230 / 240V
Power input 15A (120V), 7A (220 / 230 / 240V) 20A (120V), 10A (220V), 9A (230 / 240V)
Power supply requirements Single-phase 120V AC (50 / 60Hz) 15A or above Single-phase 120V AC (50 / 60Hz) 20A or above
Single-phase 220 / 230 / 240V AC (50 / 60Hz) 15A or above
Power Consumption (calorific value) 1.5kW (1290kcal/h) 2kW (1720kcal/h)
Accessories • Stainless steel basket with slatted bottom plate 1, • Water level sensor 1,
• Inspection sheet 1, • Warranty card 1, • Customer card 1, • Operator’s manual 1,
• Clear plastic accessory case 1, • Attachment screw for clear plastic accessory case 1
Stainless basket 1 (φ224 x 376mm) Stainless basket 2 (φ300 x 182mm)