TOMY Speedy Autoclave With Drier SDL-320

Description

Autoclaves Sliding Door Type Series- Speedy Autoclave With Drier SDL-320

 

Image result for Speedy Autoclave With Drier SDL-320

 

 

Space-saving design minimizes laboratory space occupied.

The area required for installation is reduced to approximately 0.34m2 (W D: 605mm 565mm). The sliding door type model also minimizes door clearance to preserve valuable laboratory space.

Easy-to-use control panel*The control panel lighting image shown above may vary from actual display.

A real time work monitor displays 6 status indicators for each working process of the autoclave.

Easy-to-see Control PanelWhen the autoclave is turned on, the work monitor simultaneously indicates “Main On”. After setting parameters, the monitor displays status indicators of the current working process, including “Sterilizer”, “No Water”, “Exhaust”, “Dry” and “Finish”, which provides an at-a-glance monitoring of the operation status.

*The control panel lighting image shown above may vary from actual display.When the autoclave is turned on, the work monitor

The autoclave is designed with safety features that ensure operator safety.

Boil-dry safety mechanism

  • A water level sensor is electrically responsive to the volume of water inside the chamber. If it detects low water levels, the work monitor on the control panel indicates “No Water”.

Lid open / close direction

  • Lid opening/closing mechanism reduces the risk to the operator of being exposed to hot air and steam escaping from the chamber.

Current leakage breaker

  • It quickly detects abnormalities in the electrical system and protects the operator against electric shock caused by any malfunction such as a short circuit.

Gravity-operated safety valve

  • A safety valve fitted in the front panel allows easy monitoring of its operation status. When the pressure inside the chamber is too high, the safety valve is activated to protect the operator against hazardous conditions.

Sterilizing-Drying Process

SDL-320 employs an open heat-drying system, which performs drying with open lid after sterilization.

Sterilizing-Drying Process

Specifications

Model SDL-320
Operating temperature range (during sterilizing) 60 – 132°C
Operating pressure range 0 – 186kPa
Maximum operating pressure 216kPa
Temperature control Electronic
Temperature display / display range Analog / 80 – 160°C
Pressure display / display range Analog / 0 – 400kPa
Heat source 2kW electric heater
Timer control Electrically-operated
Timer display Analog
Timer
display range
Sterilizing timer 0 – 60min. 50 / 60Hz
Drying timer 0 – 60min. 50 / 60Hz
Safety devices • Empty heating prevention, • Leakage breaker, • A dead weight safety valve
Malfunction prevention devices • Low water level detection
Leakage breaker Rated breaking current: 30A (120V) , 15A (220 / 230 / 240V)
Rated sensed current: 30mA (120V) , 10mA (220 / 230 / 240V)
Protection type Class 1 equipment
Operating environment Ambient temperature: 5 – 35°C
Relative humidity: 30 – 85%
Atmospheric pressure: 700 – 1060hPa
Gradient: Within 2°
Dimensions φ325 x 770mm
Usable dimension φ315 x 670mm
Capacity of chamber 55ℓ
Chamber material SUS304
Capacity / material of drain trap 10ℓ / polyethylene
Type/material of lid gasket Accommodating internal pressure / silicone rubber
Dimensions of main unit (mm) 605W x 565D x 1080H (Height from floor to control panel:880)
Net weight 100kg
Rated Voltage AC 120 / 220 / 230 / 240V
Power input 20A (120V), 10A (220V), 9A(230 / 240V)
Power supply requirements Single-phase 120V AC (50 / 60Hz) 20A or above
Single-phase 220 / 230 / 240V AC (50 / 60Hz) 15A or above
Power Consumption (calorific value) 2kW (1720kcal/h)
Accessories • Drain trap 1, • Drain hose 1, • Warranty card 1, • Customer card 1,
• Operator’s manual 1, • Stainless basket 2 (φ300 x 182mm)